Digital Clamp Meters Recall

Fluke is voluntarily recalling its digital clamp meters - Fluke 373, 374, 375, and 376 - that were manufactured between Sept. 1, 2010, and Oct. 31, 2012.

The reason for the recall is that the printed circuit assembly in these units may not be properly fastened to the test lead input jack. This may result in inaccurate voltage readings, including a low or no voltage reading on a circuit energized with a hazardous voltage, presenting a shock, electrocution, or thermal burn hazard.